We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

We aimed to assess the difference in the accuracy of readings from automated office blood pressure machines with each other or with manual office blood pressure measurements in Chinese individuals. We collected awake 48-h ambulatory blood pressure monitoring, two automated office blood pressure device (BpTRU and WatchBP) readings, and manual office blood pressure measurements in Chinese patients (n = 135) with hypertension in a randomized sequence. Differences were compared using paired t-tests and Bland-Altman plots. The sensitivity and specificity of the techniques for detecting elevated blood pressure were calculated using awake ambulatory blood pressure monitoring as the reference standard. The WatchBP device's and awake ambulatory blood pressure readings were similar. The BpTRU device provided significantly lower mean systolic (P 