We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

BACKGROUND: Surgical technology has led to significant improvements in patient outcomes. However, failures in equipment and technology are implicated in surgical errors and adverse events. We aim to determine the proportion and characteristics of equipment-related error in the operating room (OR) to further improve quality of care. METHODS: A systematic review of the published literature yielded 19 362 search results relating to errors and adverse events occurring in the OR, from which 124 quantitative error studies were selected for full-text review and 28 were finally selected. RESULTS: Median total errors per procedure in independently-observed prospective studies were 15.5, interquartile range (IQR) 2.0-17.8. Failures of equipment/technology accounted for a median 23.5% (IQR 15.0%-34.1%) of total error. The median number of equipment problems per procedure was 0.9 (IQR 0.3-3.6). From eight studies, subdivision of equipment failures was possible into: equipment availability (37.3%), configuration and settings (43.4%) and direct malfunctioning (33.5%). Observed error rates varied widely with study design and with type of operation: those with a greater burden of technology/equipment tended to show higher equipment-related error rates. Checklists (or similar interventions) reduced equipment error by mean 48.6% (and 60.7% in three studies using specific equipment checklists). CONCLUSIONS: Equipment-related failures form a substantial proportion of all error occurring in the OR. Those procedures that rely more heavily on technology may bear a higher proportion of equipment-related error. There is clear benefit in the use of preoperative checklist-based systems. We propose the adoption of an equipment check, which may be incorporated into the current WHO checklist.

Original publication

DOI

10.1136/bmjqs-2012-001778

Type

Journal article

Journal

BMJ Qual Saf

Publication Date

09/2013

Volume

22

Pages

710 - 718

Keywords

Checklists, Healthcare quality improvement, Patient safety, Safety culture, Surgery, Equipment Failure, Humans, Medical Errors, Operating Rooms, Patient Safety